Annual QA (TG-142)

Full TG-142 / MPPG 9 annual QA checklist -- including TG-51 photon & electron reference dosimetry -- with tolerance grading and a printable report.

Record each test result below. For measured tests, enter the deviation from the expected value — a result passes when its absolute deviation is within the tolerance shown. Tolerances are pre-filled with the selected machine class default and may be edited per test.
Annual QA — header
Dosimetry
TG-51 reference dosimetry. Run the absolute output calibration in the dedicated TG-51 calculator for each photon and electron energy, then record the result below on “X-ray / electron output calibration, TG-51” as the deviation from your expected output. Each calculator opens in a new tab, so this checklist is preserved.
Test Tolerance Result
X-ray / electron output calibration, TG-51 (absolute) — TG-142 Table III ± % %
Photon flatness change from baseline — TG-142 Table III ± % %
Photon symmetry change from baseline — TG-142 Table III ± % %
Electron flatness change from baseline — TG-142 Table III ± % %
Electron symmetry change from baseline — TG-142 Table III ± % %
Photon beam quality (PDD10 or TMR20/10) from baseline — TG-142 Table III ± % %
Electron beam quality (R50) from baseline — TG-142 Table III ± mm mm
Field-size dependent output factors, x-ray (2% for FS <4x4 cm2, 1% for >=4x4 cm2) — TG-142 Table III ± % %
Output factors for electron applicators (spot check) from baseline — TG-142 Table III ± % %
Photon MU linearity / output constancy (2% for >=5 MU; 5% for 2-4 MU) — TG-142 Table III ± % %
Electron MU linearity / output constancy (>=5 MU) — TG-142 Table III ± % %
X-ray output constancy vs dose rate, from baseline — TG-142 Table III ± % %
X-ray output constancy vs gantry angle, from baseline — TG-142 Table III ± % %
Electron output constancy vs gantry angle, from baseline — TG-142 Table III ± % %
X-ray / electron off-axis factor constancy vs gantry angle — TG-142 Table III ± % %
Physical wedge transmission factor constancy — TG-142 Table III ± % %
Arc mode (expected MU, degrees) from baseline — TG-142 Table III ± % %
SRS arc rotation -- MU set vs delivered (1 MU or 2%; SRS/SBRT machines) — TG-142 Table III ± MU MU
SRS arc rotation -- gantry arc set vs delivered (1 deg or 2%; SRS/SBRT machines) — TG-142 Table III ± deg deg
Small-field / cone relative output factors, from baseline (2% for >1 cm, 5% for <=1 cm apertures; SRS/SBRT) — MPPG 9.a Table 1 ± % %
Mechanical
Test Tolerance Result
Collimator rotation isocenter, from baseline — TG-142 Table III ± mm mm
Gantry rotation isocenter, from baseline — TG-142 Table III ± mm mm
Couch rotation isocenter, from baseline — TG-142 Table III ± mm mm
Coincidence of radiation and mechanical isocenter — TG-142 Table III ± mm mm
Table-top sag under clinical load, from baseline — TG-142 Table III ± mm mm
Table angle accuracy — TG-142 Table III ± deg deg
Table travel -- maximum range of movement, all directions — TG-142 Table III ± mm mm
Electron applicator interlocks — TG-142 Table III Pass / Action / Fail
Stereotactic accessories / lockouts (SRS/SBRT machines) — TG-142 Table III Pass / Action / Fail
End-to-end hidden-target localization test (SRS/SBRT machines) — MPPG 9.a Table 1 ± mm mm
MLC
Test Tolerance Result
MLC transmission (avg leaf & interleaf), from baseline — TG-142 Table V ± % %
MLC leaf-position repeatability — TG-142 Table V ± mm mm
MLC spoke-shot radius — TG-142 Table V ± mm mm
MLC light field vs x-ray field coincidence (all energies) — TG-142 Table V ± mm mm
Segmental (step-and-shoot) IMRT test -- max error RMS — TG-142 Table V ± mm mm
Moving-window IMRT test, four cardinal angles -- max error RMS — TG-142 Table V ± mm mm
Safety
Test Tolerance Result
Manufacturer safety / interlock test procedures — TG-142 Table III Pass / Action / Fail
Door closing safety — MPPG 8.b Table 6 Pass / Action / Fail
Emergency-off switches and emergency procedures — MPPG 8.b Table 6 Pass / Action / Fail
Imaging
Test Tolerance Result
Planar MV imaging -- full range of travel SDD — TG-142 Table VI ± mm mm
Planar MV imaging dose vs baseline — TG-142 Table VI Pass / Action / Fail
Planar kV imaging beam quality / energy vs baseline — TG-142 Table VI Pass / Action / Fail
Planar kV imaging dose vs baseline — TG-142 Table VI Pass / Action / Fail
Cone-beam CT imaging dose vs baseline — TG-142 Table VI Pass / Action / Fail
Cancel
Clinical disclaimer. The default tolerances shown are TG-142 / MPPG 8.b / MPPG 9 starting points only. They must be reviewed, set, and approved by a Qualified Medical Physicist for your institution before clinical use. Radiant QuikQA is a QA aid, not a vendor-validated QA product. All results must be independently verified.