TG-51 / TG-39 Cross-Cal

Parallel-plate chamber cross-calibration against a calibrated cylindrical Farmer in a high-energy electron beam. Returns the effective N_Dw for the PP chamber.

TG-51 / TG-39 parallel-plate cross-calibration. Both chambers are read in the same high-energy electron beam (R50 ≥ 5 cm preferred). The calibrated cylindrical Farmer fixes Dw at dref; the parallel-plate chamber's effective ND,w is back-computed from its simultaneous reading.
Identity & beam
Cylindrical (reference) chamber
Parallel-plate chamber (being calibrated)
Electrometer (shared, per ADCL cert)
Environment & voltages (shared)
Cylindrical readings (nC)

Enter M+ and M− as signed raw readings. The dref + shift reading drives the cylindrical chamber's gradient correction (Pgr,Q).

Parallel-plate readings (nC)

PP chambers have no gradient correction (flat front face), so no shifted reading is needed.

Cancel
Parallel-plate kR50,ecal data ships UNVALIDATED. Verify the resulting ND,w,pp against your trusted method before clinical use. There is no PDF for this calc yet — results display on the results page. Calc engine reused verbatim from the native Radiant QuikQA app.