TG-51 / TG-39 Cross-Cal
TG-51 / TG-39 Cross-Cal
Parallel-plate chamber cross-calibration against a calibrated cylindrical Farmer in a high-energy electron beam. Returns the effective N_Dw for the PP chamber.
TG-51 / TG-39 parallel-plate cross-calibration. Both chambers are read in the
same high-energy electron beam (R50 ≥ 5 cm preferred). The
calibrated cylindrical Farmer fixes Dw at dref; the
parallel-plate chamber's effective ND,w is back-computed from its
simultaneous reading.
Parallel-plate kR50,ecal data ships UNVALIDATED. Verify the resulting
ND,w,pp against your trusted method before clinical use. There is no
PDF for this calc yet — results display on the results page. Calc engine
reused verbatim from the native Radiant QuikQA app.